Key result
Transvenous lead extraction using bidirectional rotational mechanical sheaths in octogenarians achieved a 93.9% complete procedural success rate and 98.3% clinical success rate per lead.
Why the study?
Data on the safety and efficacy of transvenous lead extraction in octogenarians with long lead dwell times using powered extraction tools were limited.
Is transvenous lead extraction using bidirectional rotational mechanical sheaths safe and effective in octogenarians?
Cohort (n=83)
Yes
Is transvenous lead extraction using bidirectional rotational mechanical sheaths safe and effective in octogenarians?
Transvenous lead extraction using bidirectional rotational mechanical sheaths is highly effective and reasonably safe in octogenarians, suggesting age alone should not preclude the procedure despite significant comorbidity-driven mortality.
Supports TLE feasibility with rotational sheaths in selected octogenarians; extends limited data but leaves open generalizability.
BACKGROUND: Outcomes of transvenous lead extraction (TLE) are well reported in the general population, However, data on safety, efficacy of TLE in octogenarians with a long lead dwell time, using powered extraction tools are limited. The aim of this multicenter study was to evaluate the safety, effectiveness of TLE in octogenarians using the bidirectional rotational mechanical sheaths and mid-term outcome after TLE. METHODS: The study population comprised 83 patients (78.3% male; mean age 85 ± 3 years; [range 80-94 years]) with 181 target leads. All the leads (mean implant duration 112 ± 77 months [range 12-377]) were extracted exclusively using the Evolution RL sheaths (Cook Medical, Bloomington, IN, USA). RESULTS: The main indication for TLE was infection in 84.3% of cases. Complete procedural success rate, clinical success rate, per lead were 93.9% and 98.3%, respectively. Failure of lead extraction was seen in 1.7% of leads. The additional use of a snare was required in 8.4% of patients. Major complications occurred in one patient (1.2%). Thirty-day mortality after TLE was 6%. During a mean time follow-up of 22 ± 21 months, 24 patients (29%) died. No procedure-related mortality occurred. Predictors of mortality included ischemic cardiomyopathy (HR 4.35; 95% CI 1.87-10.13; p = .001), left ventricularejection fraction ≤35% (HR 7.89; 95% CI 3.20-19.48; p < .001), and TLE for systemic infection (HR 4.24; 95% CI 1.69-10.66; p = .002). CONCLUSIONS: At experienced centers bidirectional rotational mechanical sheaths combined with different mechanical tools and femoral approach allowreasonable success and safety in octogenarian with long lead dwell time. Patient's age should not influence the decision to extract or not the leads, although the 30-day and mid-term mortality are significant, especially in the present of specific comorbidities.
No takes yet. Share an insight, caveat, or question.
Migliore et al. (2023) conducted a cohort in Transvenous lead extraction in octogenarians (n=83). Transvenous lead extraction using bidirectional rotational mechanical sheaths was evaluated on Complete procedural success rate per lead. Transvenous lead extraction using bidirectional rotational mechanical sheaths in octogenarians achieved a 93.9% complete procedural success rate and 98.3% clinical success rate per lead.