Key result
Left-sided IVC is linked to 67% risk of multiple deployments during leadless pacemaker implantation.
Why the study?
While some patients face difficulty during leadless pacemaker implantation, the relationship between inferior vena cava location and procedural difficulty remained uncertain.
Does left-sided IVC location increase the risk of multiple deployments during leadless pacemaker implantation in patients with bradycardia?
Cross-Sectional (n=96)
No
Does left-sided IVC location increase the risk of multiple deployments during leadless pacemaker implantation in patients with bradycardia?
A left-sided inferior vena cava is independently associated with a higher risk of requiring multiple deployments during leadless pacemaker implantation, highlighting the utility of preprocedural CT.
Answers are grounded in this paper and cite their sources.
Preprocedural CT assessment of IVC anatomy should not yet alter leadless pacemaker implantation strategy; leaves open whether this association is clinically actionable.
Introduction While some patients faced the difficulty to implant leadless pacemaker, the relationship between inferior vena cava (IVC) location and procedural difficulty was uncertain. Methods We conducted a single‐center, cross‐sectional study including 96 consecutive patients who underwent implantation of Micra leadless pacemakers between July 2022 and November 2024. Based on the positional relationship between the IVC and the right ventricular septum assessed by preprocedural computed tomography, patients were classified into right‐sided and left‐sided IVC groups. The primary outcome was multiple deployments defined as the number of device deployments ≥ 2. Results A total of 96 patients were analyzed, including 81 in the right‐sided and 15 in the left‐sided IVC group. The mean age was 84, and 55 (57%) were male patients. Multiple deployments were observed in 39 (41%) patients. The risk of the primary outcome was significantly higher in the left‐sided IVC group than in the right‐sided (67% vs. 36%), with an adjusted odds ratio of 9.5 (95% CI: 1.5–58.7, p = 0.02). Procedure time was numerically longer in the left‐sided group (36 vs. 25 min, p = 0.11). Pericardial effusion occurred only in two patients in the right‐sided group, none requiring intervention. Conclusions The left‐sided IVC was associated with multiple deployments of a leadless pacemaker and preoperative computed tomography was useful in assessing the procedural difficulty.
No takes yet. Share an insight, caveat, or question.
Kimura et al. (2026) conducted a cross-sectional in Bradycardia requiring leadless pacemaker (n=96). Left-sided inferior vena cava vs. Right-sided inferior vena cava was evaluated on Multiple deployments defined as the number of device deployments ≥ 2 (adjusted OR 9.5, 95% CI 1.5-58.7, p=0.02). A left-sided inferior vena cava was associated with a higher risk of requiring multiple deployments during leadless pacemaker implantation compared to a right-sided IVC.
Synapse has enriched 10 related papers. Consider them for comparative context: