Key result
Long-axis measurements of myocardial T1 and ECV using MOLLI and SASHA showed no significant differences compared to short-axis measurements, demonstrating good agreement.
Why the study?
Very limited data are available comparing CMR T1 and ECV measurements obtained from short versus long axis acquisitions.
Do long axis acquisitions of T1 and ECV mapping using MOLLI and SASHA agree with short axis acquisitions in patients with myocardial infarction and healthy volunteers?
Cross-Sectional (n=16)
No
Do long axis acquisitions of T1 and ECV mapping using MOLLI and SASHA agree with short axis acquisitions in patients with myocardial infarction and healthy volunteers?
Long axis measurements of myocardial T1 and ECV using MOLLI and SASHA exhibit good agreement with short axis measurements, enabling faster myocardial tissue characterization.
Answers are grounded in this paper and cite their sources.
Slice orientation may affect T1/ECV values; leaves open standardized protocols for reproducible myocardial tissue characterization.
BACKGROUND: Although previous studies have examined the impact of slice position in volumetric measurements in Cardiovascular Magnetic Resonance (CMR) imaging, very limited data are available today comparing T1 and Extra-Cellular Volume (ECV) measurements from short and long axis acquisitions. The purpose of this study was to investigate the impact of slice position and orientation on T1 and ECV measurements using the MOdified Look-Locker Inversion recovery (MOLLI) and Saturation recovery single-shot acquisition (SASHA) sequence in patients with myocardial infarction and in healthy volunteers. METHODS: Eight (8) healthy volunteers with no medical history and eight (8) patients with myocardial infarction were included in this study. MOLLI and SASHA were utilized and short-axis and long-axis images were acquired. T1 and ECV measurements were performed by drawing same size regions of interest on the myocardium as well in the blood pool at the intersections of the short axis and long axis images. RESULTS: In healthy volunteers, there were no statistically significant differences in native T1 and ECV values between short axis and long axis acquisitions using MOLLI (two-chamber, three-chamber and four-chamber) and SASHA (three-chamber). In patients, there were no statistically significant differences in native T1 and ECV values between short axis and 3-chamber long axis acquisitions in both remote and affected myocardium using MOLLI and SASHA. CONCLUSIONS: Long axis measurements of myocardial T1 and ECV using MOLLI and SASHA exhibit good agreement with the corresponding short axis measurements allowing for fast and reliable myocardial tissue characterization in cases where shortening of the overall imaging acquisition is required.
No takes yet. Share an insight, caveat, or question.
Xanthis et al. (2019) conducted a cross-sectional in Myocardial infarction (n=16). Long-axis acquisition (MOLLI and SASHA) vs. Short-axis acquisition was evaluated on Native T1 and ECV values (p=>0.05). Long-axis measurements of myocardial T1 and ECV using MOLLI and SASHA showed no significant differences compared to short-axis measurements, demonstrating good agreement.