Key result
Digital image correlation for pulse wave velocity detection yielded values of 3.68-5.19 m/s, with a maximum absolute difference of 2.78 m/s compared to ultrasound.
Why the study?
Does digital image correlation accurately measure pulse wave velocity compared to ultrasound in healthy subjects?
Observational (n=5)
Does digital image correlation accurately measure pulse wave velocity compared to ultrasound in healthy subjects?
Digital image correlation represents a novel, feasible method for full-field time-resolved assessment of arterial stiffness and pulse wave velocity.
Answers are grounded in this paper and cite their sources.
DIC supports feasibility for PWV assessment in healthy subjects; leaves open clinical accuracy versus ultrasound.
Pulse wave velocity (PWV) of the arterial system is a very important parameter to evaluate cardiovascular health. Currently, however, there is no golden standard for PWV measurement. Digital image correlation (DIC) was used for full-field time-resolved assessment of displacement, velocity, acceleration, and strains of the skin in the neck directly above the common carotid artery. By assessing these parameters, propagation of the pulse wave could be tracked, leading to a new method for PWV detection based on DIC. The method was tested on five healthy subjects. As a means of validation, PWV was measured with ultrasound (US) as well. Measured PWV values were between 3.68 and 5.19 m/s as measured with DIC and between 5.14 and 6.58 m/s as measured with US, with a maximum absolute difference of 2.78 m/s between the two methods. DIC measurements of the neck region can serve as a test base for determining a robust strategy for PWV detection, they can serve as reference for three-dimensional fluid-structure interaction models, or they may even evolve into a screening method of their own. Moreover, full-field, time-resolved DIC can be adapted for other applications in biomechanics.
No takes yet. Share an insight, caveat, or question.
Campo et al. (2014) conducted an observational in Healthy (n=5). Digital image correlation (DIC) vs. Ultrasound (US) was evaluated on Pulse wave velocity (PWV) (maximum absolute difference 2.78 m/s). Digital image correlation for pulse wave velocity detection yielded values of 3.68-5.19 m/s, with a maximum absolute difference of 2.78 m/s compared to ultrasound.
Synapse has enriched 10 related papers. Consider them for comparative context: