Key result
Contact force sensing during atrial fibrillation ablation showed inadequate contact was most common at the ridge, while excessive contact occurred mostly around the right pulmonary vein.
Why the study?
What is the distribution of catheter-tissue contact force at different left atrial sites during conventional catheter ablation for atrial fibrillation?
Observational (n=23)
No
What is the distribution of catheter-tissue contact force at different left atrial sites during conventional catheter ablation for atrial fibrillation?
Answers are grounded in this paper and cite their sources.
Variable left atrial contact may limit transmural lesions; leaves open whether CF guidance improves AF ablation outcomes.
Objective The inability to create transmural lesions may cause atrial fibrillation recurrence after catheter ablation.The aim of the present study was to evaluate the catheter-tissue contact at different sites in left atrium with the contact force(CF) sensing catheter and the potential impact on ablation outcome. Methods Twenty-three patients with atrial fibrillation referred to Fuwai Hospital for the first-time catheter ablation from March to May, 2014 were enrolled [21 males, (53.6±10.6) years old, 15 cases of paroxysmal atrial fibrillation]. Without the contact indicating window, a single experienced operator attempted to achieve qualified contact in 15 left atrium sites (superior, middle and inferior part of the ridge between left pulmonary vein and left atrial appendage; left, middle and right part of the roof; posterior conjunction of left superior pulmonary vein and left inferior pulmonary vein; anterior, posterior conjunction of right superior pulmonary vein and right inferior pulmonary vein; superior, middle and inferior part of mitral isthmus; superior, middle and inferior part of septal isthmus) using the CF sensing catheter.The CF was recorded when qualified contact was considered according to the conventional criteria including electrogram amplitude, impedance, fluoroscopy and 3D navigation.Inadequate contact, qualified contact and excessive contact were defined when CF<10 g, 10-<40 g and ≥40 g, respectively. Results In total, 345 points were recorded in 23 patients, with the median of the CF 25 g (Q1-Q3: 10-23 g). Among them, points with inadequate, qualified and excessive contact were 85(24.6%), 241(69.9%) and 19(5.5%), respectively.Superior, middle and inferior ridge were the most common sites with inadequate contact, with median CF 7 g (Q1-Q3: 3-12 g)、8 g (Q1-Q3: 4-12 g) and 10 g (Q1-Q3: 4-22 g). Points with inadequate contact accounted for 56.5% (13/23), 52.2% (12/23) and 47.8% (11/23), respectively.In contrast, right roof, superior and middle of septal isthmus were the most common sites with excessive contact, with median CF 27 g (Q1-Q3: 17-32 g)、26 g (Q1-Q3: 13-37 g) and 23 g (Q1-Q3: 20-30 g). Points with excessive contact accounted for 17.4% (4/23), 21.7% (5/23) and 17.4% (4/23), respectively. Conclusion With the conventional ablation approach, ridge is the most common site with inadequate contact, which may explain recurrence after atrial fibrillation ablation, whereas excessive contact are more frequent at sites around right pulmonary vein, which may increase the risk of cardiac temponade.
No takes yet. Share an insight, caveat, or question.
Qiao et al. (2015) conducted an observational in Atrial fibrillation (n=23). Contact force sensing catheter was evaluated on Catheter-tissue contact (inadequate, qualified, excessive) at different left atrial sites. Contact force sensing during atrial fibrillation ablation showed inadequate contact was most common at the ridge, while excessive contact occurred mostly around the right pulmonary vein.