Key result
Internal cardioversion using a new balloon-guided single lead catheter successfully restored sinus rhythm in 92.6% (25 of 27) of patients with atrial fibrillation.
Why the study?
Does a new balloon-guided single lead catheter effectively restore sinus rhythm in patients with atrial fibrillation?
Does a new balloon-guided single lead catheter effectively restore sinus rhythm in patients with atrial fibrillation?
A new balloon-guided single lead catheter for internal cardioversion of atrial fibrillation is highly effective at restoring sinus rhythm and provides advantageous backup pacing capabilities.
Answers are grounded in this paper and cite their sources.
New IntCV device feasible in small AF cohort; leaves open efficacy and safety versus external cardioversion.
BACKGROUND: Based on the observation that internal cardioversion (IntCV) of atrial fibrillation is effective with electrodes in the right atrium and pulmonary artery, a new balloon-guided catheter and external defibrillation device with optional dual chamber pacing was evaluated. METHODS: IntCV was attempted in 27 patients (age: 57 +/- 10 years, duration: 14 +/- 18 months, left atrial diameter 56 +/- 8 mm) using a new defibrillation device (Alert, EP MedSystems, Inc., NJ, USA) that allows the delivery of biphasic shocks (0.5-15 J, variable tilt), atrial and ventricular pacing, and online signal recording. Pacing and defibrillation shocks were applied via a 7.5 Fr balloon-guided catheter (EP MedSystems, Inc.). Pacing, sensing, and triggering were established through the proximal atrial array and an electrode ring between both defibrillation arrays and a single ventricular electrode ring. Catheters were inserted from the antecubital vein. RESULTS: In 25 of 27 patients sinus rhythm was restored with a mean energy of 6.7 +/- 4.5 J. In five patients, atrial postshock pacing was required for bradycardia and atrial premature beats. The mean fluoroscopy time was 2.0 +/- 1.3 minutes. CONCLUSION: The high success rate, ease of application, and backup dual chamber pacing suggest that this system is an alternative to established methods of cardioversion. In certain indications, such as failure of prior external cardioversion and situations in which a standard pulmonary balloon catheter is needed, this system would be advantageous.
No takes yet. Share an insight, caveat, or question.
Plewan et al. (1999) studied Atrial fibrillation (n=27). Internal cardioversion using a new balloon-guided single lead catheter and external defibrillation device was evaluated on Restoration of sinus rhythm. Internal cardioversion using a new balloon-guided single lead catheter successfully restored sinus rhythm in 92.6% (25 of 27) of patients with atrial fibrillation.