Key result
Fiducial marker fitting with a parameter distribution model accurately reconstructed left ventricular deformation, with a mean difference of 0.1 ± 3.2% for circumferential strain compared to MR tagging.
Why the study?
Does fiducial marker fitting with a parameter distribution model accurately estimate left ventricular deformation compared to MR tagging in normal subjects?
Does fiducial marker fitting with a parameter distribution model accurately estimate left ventricular deformation compared to MR tagging in normal subjects?
Fiducial marker fitting combined with a parameter distribution model provides accurate reconstruction of left ventricular deformation in normal subjects, comparable to MR tagging.
Answers are grounded in this paper and cite their sources.
Supports feasibility of mesh-based LV motion estimation from 3D images; leaves open clinical validation before diagnostic adoption.
We present a method to estimate left ventricular (LV) motion based on three-dimensional (3-D) images that can be derived from any anatomical tomographic or 3-D modality, such as echocardiography, computed tomography, or magnetic resonance imaging. A finite element mesh of the LV was constructed to fit the geometry of the wall. The mesh was deformed by optimizing the nodal parameters to the motion of a sparse number of fiducial markers that were manually tracked in the images through the cardiac cycle. A parameter distribution model (PDM) of LV deformations was obtained from a database of MR tagging studies. This was used to filter the calculated deformation and incorporate a priori information on likely motions. The estimated deformation obtained from 13 normal untagged studies was compared with the deformation obtained from MR tagging. The end systolic (ES) circumferential and longitudinal strain values matched well with a mean difference of 0.1 +/- 3.2% and 0.3 +/- 3.0%, respectively. The calculated apex-base twist angle at ES had a mean difference of 1.0 +/- 2.3 degrees. We conclude that fiducial marker fitting in conjunction with a PDM provides accurate reconstruction of LV deformation in normal subjects.
No takes yet. Share an insight, caveat, or question.
Remme et al. (2005) studied Normal left ventricular function (n=13). Fiducial marker fitting with a parameter distribution model (PDM) vs. MR tagging was evaluated on End systolic (ES) circumferential and longitudinal strain values (Mean difference 0.1 ± 3.2% (circumferential) and 0.3 ± 3.0% (longitudinal)). Fiducial marker fitting with a parameter distribution model accurately reconstructed left ventricular deformation, with a mean difference of 0.1 ± 3.2% for circumferential strain compared to MR tagging.