Key result
Underfilling blood collection tubes prolongs PT and APTT results, though 3.2% sodium citrate tubes tolerate lower fill volumes than 3.8% citrate tubes.
Why the study?
Does underfilling of blood collection tubes with 3.2% vs 3.8% sodium citrate affect the results of routine coagulation assays in healthy persons and patients on oral anticoagulants?
Observational
Does underfilling of blood collection tubes with 3.2% vs 3.8% sodium citrate affect the results of routine coagulation assays in healthy persons and patients on oral anticoagulants?
Using 3.2% sodium citrate is recommended for coagulation testing because it is less sensitive to underfilling compared to 3.8% citrate, allowing acceptable results at 60% (PT) and 70% (APTT) of optimum fill volume.
Answers are grounded in this paper and cite their sources.
Underfilled citrate tubes may distort PT/APTT results; leaves open precise volume thresholds for reliable testing.
We evaluated the effect of sample volume and citrate concentration on results of routine coagulation assays (prothrombin time [PT] and activated partial thromboplastin time [APTT]). The study was performed on samples obtained from healthy persons and patients receiving oral anticoagulant therapy. Standard evacuated tubes (3.2% and 3.8% sodium citrate) were filled to varying total sample volumes ranging from 3.0 to 5.0 mL, and results of routine coagulation tests were compared. Underfilling may significantly affect the APTT and PT, resulting in artifactual prolongation of results. This effect is most pronounced in samples drawn into 3.8% citrate. By using 3.8% citrate, there is a statistically significant difference in the results of PT assays in the samples less than 80% filled compared with those that are 100% filled. For APTT assays performed on samples drawn into 3.8% citrate, a statistical difference occurred at less than 90% filled. This effect was less pronounced when samples were drawn into 3.2% sodium citrate. We found no statistically significant difference in PT results from a 3.2% citrate tube between fill volumes of 60% and 100% and none for APTT results between fill volumes of 70% and 100%. This study further supports the recommendation to use 3.2% sodium citrate concentration, because 60% of the optimum filled volume for PT and 70% of the optimum filled volume for APTT are acceptable.
No takes yet. Share an insight, caveat, or question.
Adcock et al. (1998) conducted an observational in Healthy persons and patients receiving oral anticoagulant therapy. Underfilling of standard evacuated tubes (3.2% and 3.8% sodium citrate) vs. 100% filled tubes was evaluated on Prothrombin time (PT) and activated partial thromboplastin time (APTT). Underfilling blood collection tubes prolongs PT and APTT results, though 3.2% sodium citrate tubes tolerate lower fill volumes than 3.8% citrate tubes.
Synapse has enriched 10 related papers. Consider them for comparative context: