Key result
Underfilling citrate tubes below 90% capacity yields falsely elevated PT and APTT results.
Why the study?
Underfilling of specimen tubes containing 129 mmol/L (3.8%) buffered citrate prolongs PT and APTT values, but the effect with 109 mmol/L (3.2%) buffered citrate was unclear.
Does underfilling specimen tubes with 3.2% citrate anticoagulant affect the accuracy of PT and APTT values in healthy subjects and patients on warfarin?
Does underfilling specimen tubes with 3.2% citrate anticoagulant affect the accuracy of PT and APTT values in healthy subjects and patients on warfarin?
Specimen tubes containing 3.2% citrate must be filled to at least 90% of capacity to avoid falsely elevated PT or APTT results, particularly for patients on warfarin therapy.
Answers are grounded in this paper and cite their sources.
Caution against underfilled 3.2% citrate tubes for PT/APTT; extends 3.8% data but leaves optimal fill thresholds open.
Underfilling of specimen tubes containing 129 mmol/L (3.8%) buffered citrate prolongs prothrombin time (PT) and activated partial thromboplastin time (APTT) values. We studied this phenomenon by using 109 mmol/L (3.2%) buffered citrate as the anticoagulant, anticipating some increase in tolerance to underfilling. Venous blood drawn from 12 healthy subjects and 30 patients receiving long-term oral warfarin therapy was mixed with 109 mmol/L buffered citrate solution in proportions equivalent to filling the collection tubes from 52% to 100% of capacity. Accurate PT values were obtained from normal specimens if the tubes were filled to 65% or more of capacity. Accurate PT results in the therapeutic range were obtained only with filling to 80% or more of capacity (using a "moderately sensitive" thromboplastin reagent, International Sensitivity Index [ISI] = 2.06) or 90% or more of capacity (using a "highly sensitive" thromboplastin reagent, ISI = 1.01). In contrast, APTT was much less tolerant to underfilling, with prolonged values observed in most specimens filled to less than 90% of capacity. No false low values were observed. Specimen tubes should be filled to at least 90% of capacity to avoid falsely elevated PT or APTT results, but values within the reference range may be acceptable even from underfilled tubes.
No takes yet. Share an insight, caveat, or question.
Reneke et al. (1998) studied Healthy subjects and patients on warfarin therapy (n=42). Underfilling of specimen tubes with 109 mmol/L (3.2%) citrate vs. Adequately filled tubes was evaluated on Accuracy of prothrombin time (PT) and activated partial thromboplastin time (APTT). Specimen tubes with 3.2% citrate must be filled to at least 90% of capacity to avoid falsely elevated PT or APTT results, though normal specimens may yield accurate PTs at 65% capacity.