Key result
A novel matched filter method for CW-Doppler radar achieved absolute heart rate errors of 1.93±1.76 bpm in supine and 9.72±7.86 bpm in seated postures compared to references.
Why the study?
CW-Doppler radar measures chest surface movement comprising superimposed cardiac and respiratory signals, making the extraction of cardiac information challenging.
Does a novel matched filter-based method improve the accuracy of heartbeat detection from continuous wave Doppler radar sensors in different postures?
Does a novel matched filter-based method improve the accuracy of heartbeat detection from continuous wave Doppler radar sensors in different postures?
A novel matched filter-based method using continuous wave Doppler radar can detect heart rate with low error, particularly in the supine posture, offering potential for non-contact vital sign monitoring.
May enable non-contact HR monitoring in controlled settings; hypothesis-generating and requires prospective clinical validation.
Heart rate measurement using a continuous wave Doppler radar sensor (CW-DRS) has been applied to cases where non-contact detection is required, such as the monitoring of vital signs in home healthcare. However, as a CW-DRS measures the speed of movement of the chest surface, which comprises cardiac and respiratory signals by body motion, extracting cardiac information from the superimposed signal is difficult. Therefore, it is challenging to extract cardiac information from superimposed signals. Herein, we propose a novel method based on a matched filter to solve this problem. The method comprises two processes: adaptive generation of a template via singular value decomposition of a trajectory matrix formed from the measurement signals, and reconstruction by convolution of the generated template and measurement signals. The method is validated using a dataset obtained in two different experiments, i.e., experiments involving supine and seated subject postures. Absolute errors in heart rate and standard deviation of heartbeat interval with references were calculated as 1.93±1.76bpm and 57.0±28.1s for the lying posture, and 9.72±7.86bpm and 81.3±24.3s for the sitting posture.
No takes yet. Share an insight, caveat, or question.
Iwata et al. (2021) studied Heart rate measurement. Matched filter method using singular value decomposition for CW-Doppler radar vs. Reference measurements was evaluated on Absolute errors in heart rate and standard deviation of heartbeat interval. A novel matched filter method for CW-Doppler radar achieved absolute heart rate errors of 1.93±1.76 bpm in supine and 9.72±7.86 bpm in seated postures compared to references.