The study was designed to validate the TM-2441 ambulatory blood pressure monitoring device using the ISO 81060-2:2013 standard.
Does the TM-2441 ABPM device accurately measure blood pressure compared to standard methods in healthy individuals?
Does the TM-2441 ABPM device accurately measure blood pressure compared to standard methods in healthy individuals?
The TM-2441 ABPM device is accurate and fulfills ISO 81060-2:2013 standard requirements, making it suitable for clinical use in adults with hypertension.
Answers are grounded in this paper and cite their sources.
TM-2441 ABPM meets ISO criteria in healthy volunteers; leaves open accuracy in hypertensive or clinical populations.
OBJECTIVE: The aim of this study was to validate the TM-2441 ambulatory blood pressure monitoring (ABPM) device using the ISO 81060-2:2013 standard. PARTICIPANTS AND METHODS: Participants were healthy individuals aged more than 12 years who were recruited from among the outpatients and volunteers of Jichi Medical University (Tochigi, Japan). The same-arm sequence protocol (clinical validation) and the opposite-limb simultaneous method (ambulatory validation) from the ISO 81060-2:2013 standard were used. RESULTS: One hundred and seven participants were enrolled; 85 participated in the clinical validation and 35 participated in the ambulatory validation (13 participants were included in both validation protocols). The TM-2441 device performed well against the standard in both the clinical and ambulatory validations; the mean and SD values for the differences between device and observed systolic and diastolic blood pressure values in both tests fulfilled criterion 1 and criterion 2 of the standard. The Bland-Altman plots did not show any systematic variation in the error. CONCLUSION: The TM-2441 ABPM device was accurate and fulfilled all ISO 81060-2:2013 standard requirements for ABPM determination in adults. It is therefore suitable for use for ABPM in adults with hypertension.
No takes yet. Share an insight, caveat, or question.
Kario et al. (2018) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: