Does the A&D BP UA-1200BLE device accurately measure blood pressure compared to standard measurement in individuals aged >12 years?
Does the A&D BP UA-1200BLE device accurately measure blood pressure compared to standard measurement in individuals aged >12 years?
The UA-1200BLE device is highly accurate and meets ISO 81060-2:2013 standards, making it suitable for wireless home blood pressure monitoring in patients with hypertension.
Answers are grounded in this paper and cite their sources.
Device meets ISO accuracy criteria; supports home BP monitoring in >12 years but leaves open outcome validation.
OBJECTIVE: The aim of this study was to validate the UA-1200BLE device using the ISO 81060-2:2013 standard. MATERIALS AND METHODS: Individuals aged more than 12 years were recruited from among the outpatients and volunteers of Jichi Medical University (Tochigi, Japan). Eighty-five patients were included in the final analyses. Validation was performed using the same-arm sequence protocol as per the ISO 81060-2:2013 standard. RESULTS: The device performed well against the standard; the mean and SD values for the differences between device and observed systolic and diastolic blood pressure values fulfilled both criterion 1 and criterion 2 of the standard (criterion 1: -2.62±6.89 and -0.90±6.63 mmHg, respectively; criterion 2: -2.65±5.36 and -0.91±6.12 mmHg, respectively). The Bland-Altman plots did not show any systematic variation in the error. CONCLUSION: The UA-1200BLE device had a high level of accuracy and fulfilled all ISO 81060-2:2013 standard requirements in adults with a wide range of arm circumference measurements. It is therefore suitable as a wireless blood pressure monitoring tool in patients with hypertension.
No takes yet. Share an insight, caveat, or question.
Kario et al. (2018) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: