Why the study?
The study aimed to validate the blood pressure measurement performance of the A&D UA-1100NFC hoseless devices across two cuff sizes according to the ISO 81060-2:2018/Amd 1:2020 protocol.
Do the A&D UA-1100NFC and UA-1100NFC-W hoseless upper arm-type home blood pressure devices provide accurate blood pressure measurements compared to observer-measured values?
Do the A&D UA-1100NFC and UA-1100NFC-W hoseless upper arm-type home blood pressure devices provide accurate blood pressure measurements compared to observer-measured values?
The A&D UA-1100NFC and UA-1100NFC-W hoseless upper arm-type home blood pressure devices are accurate and fulfill the ISO81060-2:2018/Amd 1:2020 validation standard, making them suitable for home BP monitoring.
Answers are grounded in this paper and cite their sources.
Supports UA-1100NFC accuracy per ISO protocol; leaves open outcome trials before clinical adoption.
The aim of this study was to validate the performance of the A&D UA-1100NFC hoseless devices of two cuff sizes in monitoring blood pressure (BP) in the upper arm according to the International Organization for Standardization (ISO) 81060-2:2018/amendment (Amd) 1:2020 protocol. The accuracy of the UA-1100NFC (for arm circumferences of 22.0-32.0 cm) and the UA-1100NFC-W (for arm circumferences of 22.0-42.0 cm) was assessed using the same-arm sequence protocol as per the ISO 81060-2:2018/Amd 1:2020 standard. Individuals aged >12 years were recruited from the outpatients and volunteers of Jichi Medical University (Tochigi, Japan); 85 participants were included in the final analyses for each device. Both devices performed well against the standard; mean and SD values for the differences between the device-measured and observer-measured SBP and DBP values met both criterion 1 and criterion 2 of the standard [(UA-1100NFC) criterion 1: -3.71 ± 6.82 mmHg and 0.86 ± 6.33 mmHg, respectively; criterion 2: 5.65 mmHg and 5.87 mmHg, respectively; (UA-1100NFC-W) criterion 1: 0.73 ± 7.84 mmHg and 1.72 ± 6.44 mmHg, respectively; criterion 2: 6.49 mmHg and 5.86 mmHg, respectively]. The Bland-Altman plots did not show any systematic variation in the error. Both the UA-1100NFC and UA-1100NFC-W hoseless devices had a high level of accuracy and fulfilled the requirements of the ISO81060-2:2018/Amd 1:2020 validation standard. They are therefore suitable as home BP monitoring tools in patients with hypertension.
No takes yet. Share an insight, caveat, or question.
Kario et al. (2025) studied this question.