Clarifies that 'pseudo-malfunction' in biventricular systems is a normal device behavior that can mimic malfunction, and discusses management of T-wave oversensing during LV pacing.
We are grateful to Bertini et al. for their interest in our paper and for their excellent comments and observations. The term ‘pseudo-malfunction’ can indeed be misleading, as it can raise the doubt that the device is not working properly. We do agree that the particular behaviour in our case is absolutely ‘normal’ and does not imply any kind of problem with the device. Moreover, it is not manufacturer-specific: it can take place with any biventricular system under the circumstances described in the paper.1 In our intentions, ‘pseudo-malfunction’ is a term used to describe a particular device feature that can ‘mimic’ a malfunction and may lead to difficulties in ECG and telemetry interpretation. Occasionally, a ‘normal’ device behaviour can mimic ventricular oversensing, ventricular undersensing, loss of capture, or can cause deviations from expected pacing rate. Failure to recognize these features can lead to costly and fruitless investigations into (non-existent) hardware malfunctions.2 This is not a clinical problem when it happens only during threshold test. On the contrary, it may become a relevant problem when left ventricular (LV) pacing alone is chosen in cardiac resynchronization therapy defibrillator devices3–5 and sensing of the T-wave is detected as a VF interval. In this particular situation, only two solutions are available (as properly stated by Bertini et al.): (i) to reduce ventricular sensitivity and (ii) to prolong ventricular refractoriness. The first choice is feasible in a defibrillator as far as it does not impair arrhythmia detection: if a low sensitivity is needed (>0.9 mV) testing of appropriate VF detection may be advisable; the second solution can work in a cardiac resynchronization therapy pacemaker device. An automatic prolongation of refractoriness or decrease of sensitivity to 3 mV during manual LV threshold testing would avoid such T-wave oversensing. In our case the third possibility was applicable, once you know that there is no malfunction: do nothing … … in the end you do not need a ‘solution’ when there is ‘no problem’! Conflict of interest: none declared.
No takes yet. Share an insight, caveat, or question.
Maria et al. (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: