Does VEC cine MR imaging demonstrate differences in blood flow velocity between the true and false channels in patients with type B aortic dissection?
Does VEC cine MR imaging demonstrate differences in blood flow velocity between the true and false channels in patients with type B aortic dissection?
VEC cine MR imaging can noninvasively quantify and demonstrate significant hemodynamic differences between the true and false channels in chronic type B aortic dissection.
Answers are grounded in this paper and cite their sources.
Supports noninvasive flow mapping in type B dissection; leaves open clinical utility pending larger validation.
Velocity encoded (VEC) cine MR imaging is a new noninvasive technique for the quantification of blood flow velocity in the cardiovascular system. Six patients with type B aortic dissection underwent VEC cine MR imaging at 1.5 T. This technique provides cine MR magnitude and VEC phase images at approximately 16 equally spaced intervals during an average cardiac cycle. A region of interest encompassing a vascular structure, i.e., false channel, provides a spatially averaged velocity for the time interval at which the image was acquired. Interpretation of velocity values from the 16 intervals during the cardiac cycle provides a temporally average velocity. Velocity mapping across the aortic lumen in these six cases showed average spatial and temporal velocity of 13.4 +/- 1.49 cm/s in the true channel and 3.1 +/- 0.84 cm/s in the false channel (p less than 0.05). The peak systolic velocity (temporal peak) was 43.6 +/- 7.20 cm/s in the true channel and 14.3 +/- 2.30 cm/s in the false channel (p less than 0.05). The flow volume per cardiac cycle was not significantly different between the ture (23.1 +/- 5.04 ml/cycle) and false channel (27.1 +/- 10.14 ml/cycle). There was substantial retrograde flow in the false channel of two patients. The intraobserver and interobserver variability was less than 10% (r = 0.98 to 0.99) for the measurement of flow parameters in both the true and the false channel. We conclude that VEC cine MR imaging demonstrates substantial differences in the hemodynamic pattern in the true and false channel in aortic dissection.
No takes yet. Share an insight, caveat, or question.
Chang et al. (1991) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: