Ultrasonic relative pressure imaging can accurately measure intraventricular pressure differences, though higher frame rates (>60 Hz) are needed to prevent underestimation of peak values.
Answers are grounded in this paper and cite their sources.
May enable noninvasive LV pressure mapping in echocardiography; leaves open prospective clinical validation.
Ultrasonic relative pressure imaging (RPI) visualizing pressure distribution on the basis of vector flow mapping (VFM) was developed by applying the momentum equations of fluid motion (Navier–Stokes equations; NSE) for the two-dimensional (2D)-VFM velocity fields. Numerical investigations using three-dimensional (3D) jets with a size similar to the left ventricle (LV) quantified degradations caused by the flow three-dimensionality of convection in the range of 0.1 mmHg. Experimental validation using an LV phantom quantitatively compared two-point intraventricular pressure differences (IVPDs) obtained by RPI, two pressure sensors (grand truth), and color M-mode (CMM) with a high frame rate of 167 Hz in diastole where a transient term is dominant. All measurements of IVPDs in diastole were in good agreement, although the peak value of IVPD by RPI was underestimated by about 35%, mainly because of its insufficient frame rate of 35 Hz. A dependence of the frame rate on IVPD was examined by down-sampled CMM. Although IVPD based on down-sampled CMM at 27 Hz similar to that of the current RPI frame rate showed that the accuracy was underestimated by 30%, a higher data rate of more than 60 Hz, which could be realized in the near future, provided reasonable accuracy with less than 10% error.
No takes yet. Share an insight, caveat, or question.
Tanaka et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: