High concentrations of calcium indicators like Fura-2 or Indo-1 may artificially alter the physiological calcium transients and action potentials they are meant to measure in cardiac muscle.
Answers are grounded in this paper and cite their sources.
High probe levels may artifactually alter cardiac Ca2+ transients and APs in experiments; leaves open optimal indicator dosing for accurate measurements.
The use of high-affinity fluorescent probes for monitoring intracellular free Ca2+ in cardiac muscle is now widespread. We have investigated the consequences of introducing intracellular buffers with the properties of Fura-2 or Indo-1 on the action potential, Ca2+ transient and contractile activity of the myocardium. Our theoretical results suggest that, at the high intracellular concentrations of these fluorescent probes used on occasion to improve the signal-to-noise ratio of the emitted fluorescence, modulation of action potential profile and attenuation of the amplitudes of the Ca2+ transient and contraction can occur, together with subtle changes in the kinetics of these events.
No takes yet. Share an insight, caveat, or question.
Noble et al. (1991) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: