Why the study?
Does ICD lead implantation in the RVOT provide comparable defibrillation thresholds and pacing/sensing characteristics to RVA implantation in patients requiring ICDs?
Does ICD lead implantation in the RVOT provide comparable defibrillation thresholds and pacing/sensing characteristics to RVA implantation in patients requiring ICDs?
ICD lead placement in the RVOT is a viable alternative to the RVA with comparable defibrillation thresholds and sensing/pacing characteristics, though active-fixation leads are recommended to prevent dislodgment.
Answers are grounded in this paper and cite their sources.
Comparable DFTs with RVOT versus RVA leads; hypothesis-generating and should not yet change practice.
Unipolar ICD electrodes are routinely implanted at the right ventricular apex (RVA). However, inappropriate pacing/sensing parameters and/or high DFT may limit the appropriateness of the lead's implantation at the RVA. This study examined the effects on DFT of ICD leads implanted in the RVOT, attached to the high interventricular septum as an alternate location. DFT, defibrillation impedance, and sensing and pacing characteristics were measured at the time of implantation in 28 consecutive patients. Group A consisted of 12 patients in whom the ICD implantation criteria in the RVA were not satisfied, and whose lead was placed in the RVOT. Group B consisted of 16 patients with ICD electrodes implanted at the RVA. Mean DFT in group A was 11 +/- 4 J (4.5-20 J) versus 12 +/- 6 J (4-20 J) in the group B (P = 0.58). Defibrillation impedance was 81 +/- 9 omega (69-92 omega) in group A versus 77 +/- 15 omega) (46-93 omega) in group B (P = 0.43). R wave amplitude, slew rate, pacing threshold, and pacing impedance were comparable in both groups. In the perioperative period, the electrode needed to be repositioned in two patients from group A. There was no further dislodgment of RVOT defibrillation leads or other lead related complications during a follow-up of 23 +/- 9 months. The placement of ICD leads in the RVOT is an alternative to the RVA position. However, active-fixation ICD leads should be considered to limit the risk of electrode dislodgment.
No takes yet. Share an insight, caveat, or question.
Lubiński et al. (2000) studied this question.