Key result
Novel ICD biphasic waveform system lowers defibrillation thresholds by ~24% at 3 months.
Why the study?
The optimal pulse durations and tilts of biphasic defibrillation waveforms for transvenous ICD systems have yet to be identified.
p-value: p=0.03
New biphasic waveform with steroid-eluting RV lead yielded low stable DFTs; supports hypothesis-generating evaluation before practice change.
Biphasic defibrillation waveforms have provided a reduction in defibrillation thresholds in transvenous ICD systems. Although a variety of biphasic waveforms have been tested, the optimal pulse durations and tilts have yet to be identified. A multicenter clinical study was conducted to evaluate the performance of a new ICD biphasic waveform and new RV active fixation steroid eluting lead system. Fifty-three patients were entered into the study. Mean age was 63 years with a mean ejection fraction of 36.8%. Primary indication for implantation was monomorphic ventricular tachycardia alone (54.7%). Forty-eight patients (90.6%) were implanted with an RV shocking lead and active can alone as the anodal contact. The ICD can was the cathode. In four cases (7.5%), an additional SVC or CS lead was used due to a high DFT with the RV lead alone. In an additional case, a chronic SVC lead was used although the RV-Can DFT was acceptable. DFT for all cases at implant was 9.8 +/- 3.7 J. Repeat testing at 3 months for a subset of patients showed a reduction in DFT (7.4 +/- 3.0 J), P value = 0.03. Sensing and pacing characteristics of the RV lead system remained excellent during the study period (acute 0.047 +/- 0.005 ms at 5.4 V and 9.9 +/- 6.2 mV R wave; chronic 0.067 +/- 0.11 ms at 5.4 V and 9.3 +/- 5.4 mV R wave). It is concluded that this lead system provides good acute and chronic sensing and pacing characteristics with good DFT values in combination with this waveform.
No takes yet. Share an insight, caveat, or question.
MEHDIRAD et al. (1999) studied Implantable cardioverter-defibrillator indication (n=53). New ICD biphasic waveform and RV active fixation steroid eluting lead system was evaluated on Defibrillation threshold (DFT) and sensing/pacing characteristics (p=0.03). A new ICD biphasic waveform and RV lead system provided good sensing and pacing characteristics with a defibrillation threshold of 9.8 ± 3.7 J at implant, decreasing to 7.4 ± 3.0 J at 3 months (P=0.03).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: