Key result
The 99th percentiles for troponin I were 21 ng/L for ARCHITECT STAT, 31 ng/L for ADVIA Centaur, and 28 ng/L for Dimension Vista, with no age dependency demonstrated.
Cross-Sectional (n=2,404)
Quantile regression established assay-specific 99th percentiles for three contemporary troponin I assays, highlighting the lack of age dependency and variable gender differences across assays.
Supports assay-specific troponin I thresholds without age adjustment; leaves open gender standardization and cross-platform harmonization.
BACKGROUND: The IFCC Task Force on Clinical Applications of Cardiac Biomarkers suggests comparing several contemporary sensitive troponin assays in the same, sufficiently large reference population. METHODS: Three contemporary sensitive assays were used to measure troponin I concentration in samples from a uniquely large healthy population (2404 individuals) and in a sub-group with tighter inclusion criteria of 908 individuals. The 99th percentiles were calculated using quantile regression which takes the entire population into account. RESULTS: The 99th percentile for the ARCHITECT STAT Troponin I assay was 21 ng/L, 31 ng/L for the ADVIA Centaur Troponin I-Ultra assay and 28 ng/L for the Dimension Vista cTnI assay. Significantly higher values were found in males than in women only in the Dimension Vista cTnI assay and in the subgroup for the ARCHITECT STAT Troponin I assay. CONCLUSIONS: Quantile regression provides a tool to accurately estimate the 99th percentile and establish a continuous function of the relation between the 99th percentile and the age and gender. There was no age dependency demonstrated. A gender difference was found in one assay but inconclusive in another and not demonstrated in a third.
No takes yet. Share an insight, caveat, or question.
Petersmann et al. (2013) conducted a cross-sectional in Healthy (n=2,404). Contemporary sensitive troponin I assays was evaluated on 99th percentile of troponin I concentration. The 99th percentiles for troponin I were 21 ng/L for ARCHITECT STAT, 31 ng/L for ADVIA Centaur, and 28 ng/L for Dimension Vista, with no age dependency demonstrated.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: