Why the study?
Do noninvasive central blood pressure devices (SphygmoCor and Omron) accurately estimate central blood pressure compared to invasive catheter measurement?
Do noninvasive central blood pressure devices (SphygmoCor and Omron) accurately estimate central blood pressure compared to invasive catheter measurement?
SphygmoCor and Omron devices closely correlate with invasive central BP measurements but underestimate absolute values, highlighting the need for device-specific diagnostic thresholds.
Noninvasive central SBP devices underestimate invasive values despite high correlation; supports device-specific thresholds but leaves open prospective outcome validation.
BACKGROUND: To investigate the accuracy of the SphygmoCor and Omron HEM9000-AI devices in the estimation of central blood pressure (BP) in comparison with the simultaneous invasive catheter measurement. METHODS: The radial arterial pulse was sequentially recorded by the use of the Omron and SphygmoCor devices in 33 patients, with the calibration of the brachial oscillometric BP, to derive central BPs, which were also measured simultaneously with a catheter-based fluid-filled manometer system. The procedure was repeated three times to obtain 99 pairs of noninvasive and invasive measurements. RESULTS: The noninvasive central systolic BP estimations were significantly (P < 0.001) associated with the invasive measurement at the ascending aorta, with a correlation coefficient of 0.91 and 0.90 for the SphygmoCor and Omron devices, respectively. However, both devices underestimated central systolic BP with a difference of -15 mm Hg (95% confidence interval (CI), -17 to -13 mm Hg, P < 0.001) for SphygmoCor and -2 mm Hg (95% CI, -4 to 0 mm Hg, P < 0.05) for Omron. In comparison with the invasive catheter measurement at the brachial artery, the oscillometric Omron device underestimated brachial systolic BP by -19 mm Hg (95% CI, -23 to -15 mm Hg, P < 0.001). CONCLUSIONS: Both devices underestimated central systolic BP, with a larger deviation by SphygmoCor. Nonetheless, these noninvasive estimations of central BP closely correlate with the invasive measurements, and can still be properly used, on the condition that device specific diagnostic thresholds become available.
No takes yet. Share an insight, caveat, or question.
Ding et al. (2011) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: