Why the study?
Does the Omron HEM-9000AI provide comparable estimates of central systolic blood pressure and peripheral augmentation index to the SphygmoCor system?
Does the Omron HEM-9000AI provide comparable estimates of central systolic blood pressure and peripheral augmentation index to the SphygmoCor system?
While the Omron HEM-9000AI and SphygmoCor systems show highly correlated measurements for pSBP2 and pAIx, the Omron system yields significantly higher estimates of central systolic blood pressure, highlighting the need for invasive validation.
Answers are grounded in this paper and cite their sources.
Omron and SphygmoCor cSBP estimates are not interchangeable; leaves open which better approximates invasive central pressure.
Central blood pressure is increasingly considered to be more predictive of cardiovascular events than peripheral pressure.Central systolic pressure (cSBP) can be estimated from the late systolic shoulder (pSBP 2 ) of the radial pulse waveform.The Omron HEM-9000AI utilizes this technique to non-invasively measure cSBP through semi-automated applanation tonometry.The aim of this study was to compare the Omron HEM-9000AI with an existing device, the SphygmoCor, and to test its reproducibility.33 subjects were tested using both the Omron and the SphygmoCor devices, which identify pSBP 2 in the radial waveform, and estimate cSBP and peripheral augmentation index (pAIx) using their inbuilt software.10 subjects were retested after a period of 24 h to determine reproducibility.There was a strong correlation (r Z 0.94; P < 0.001), and no difference between mean values of pSBP 2 (0.8 AE 4.8 mmHg; P Z 0.4) measured by the two devices.However estimates of cSBP were significantly higher in the Omron than the SphygmoCor (12.2 AE 4.6 mmHg; P < 0.001), despite significant correlation (r Z 0.95; P < 0.001).pAIx measurements were correlated (r Z 0.85; P < 0.001) with no significant difference between absolute values (mean difference Z 1 AE 9.2%; P Z 0.6).Results from both devices were reproducible, with significant correlation and insignificant mean differences for values of cSBP and pAIx.pSBP 2 values were highly correlated, but a direct analysis with invasive measurements is necessary to determine which estimate of cSBP is most accurate.Excluding this discrepancy, the Omron system is potentially useful for measuring pSBP 2 and pAIx in large scale trials.
No takes yet. Share an insight, caveat, or question.
Richardson et al. (2009) studied this question.