Why the study?
Cardiac and respiratory motion challenge accurate dose delivery during stereotactic arrhythmia radioablation (STAR).
Does a template-matching approach using kilovoltage projections accurately track ICD lead tip motion during stereotactic arrhythmia radioablation?
Does a template-matching approach using kilovoltage projections accurately track ICD lead tip motion during stereotactic arrhythmia radioablation?
A novel template-matching approach using kilovoltage projections can successfully track ICD lead tip motion in 3D during stereotactic arrhythmia radioablation, with acceptable accuracy in the majority of estimations.
May support kV-based ICD lead tracking for STAR motion management; leaves open prospective validation before clinical adoption.
Background and purpose: Cardiac and respiratory motion present a challenge to accurate dose delivery during stereotactic arrhythmia radioablation (STAR). This study investigated a motion tracking method using kilovoltage (kV) images on a gantry-mounted linear accelerator during STAR. Materials and methods: An implantable cardioverter-defibrillator (ICD) lead tip was used for motion tracking. Planning computed tomography and cone beam computed tomography scans were analysed retrospectively on ten patients treated with STAR in RAdiosurgery for VENtricular TAchycardia (RAVENTA) (four patients underwent deep inspiratory breath-hold, two with free breathing, three under abdominal compression, and one using high frequency ventilation). A template-matching approach was developed to segment ICD lead tips in kV projections, with successful segmentation defined as a segmented position within 2 mm of the two-dimensional (2D) ground truth derived from manual annotations. A probability distribution function method was used to convert 2D coordinates to three-dimensional (3D) trajectories. 3D trajectories from the 2D ground truth were used to measure ICD lead tip displacements from its mean position. Differences between 3D trajectories from 2D ground truth and segmentations were analysed. Results: The median (range) 2D segmentation success rate was 85% (71-96%). Median (range) motion displacements were 1.4 (0.1-7.4) mm (left-right), 1.5 (0.1-10.4) mm (superior-inferior), and 0.8 (0.0-6.3) mm (anterior-posterior). The median (range) proportion of 3D estimations with errors below 2 mm in all axes was 78% (43-96%). Conclusions: ICD lead tip motion showed patient-specific variability. This study presented a potential 3D motion tracking method for ICD lead tips during STAR.
No takes yet. Share an insight, caveat, or question.
Xiong et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: