Why the study?
Can the PFA-100 system detect defects in platelet adhesion or aggregation in vitro?
Can the PFA-100 system detect defects in platelet adhesion or aggregation in vitro?
The PFA-100 system can rapidly detect defects in platelet adhesion and aggregation in vitro, making it a potentially useful tool for screening platelet dysfunction.
Answers are grounded in this paper and cite their sources.
PFA-100 detects specific platelet defects in vitro; leaves open clinical screening utility pending human validation.
A new in vitro system for the evaluation of platelet function, PFA-100 (currently under evaluation) is characterized. The system monitors platelet aggrega tion on a collagen-epinephrine-coated membrane as whole blood sample is aspirated under controlled flow conditions through a microscopic aperture cut into the membrane. The time required for the platelet plug to oc clude the aperture (closure time) is indicative of the plate let function in the sample. Incubation of normal blood samples with antibodies to glycoprotein (GP) Ib, GPIIb- IIIa, von Willebrand factor, or with the peptide Arg-Gly- Asp-Ser, resulted in a concentration-dependent prolonga tion in closure time. An anti-fibrinogen antibody did not impact the closure time. Closure time was also prolonged when the hematocrit or the platelet count was lowered to pathological values. The study indicates that PFA-100 could detect defects in platelet adhesion or aggregation. The simplicity of the test makes PFA-100 unique for rapid screening of a variety of platelet dysfunction.
No takes yet. Share an insight, caveat, or question.
Kundu et al. (1996) studied this question.