Why the study?
While pacemaker reuse is well established, the safety of reusing implantable cardioverter-defibrillators regarding functionality and infection rates had not been properly investigated.
Does the use of resterilized, reused ICDs increase the rate of complications compared to new ICDs in patients with class I indications?
Does the use of resterilized, reused ICDs increase the rate of complications compared to new ICDs in patients with class I indications?
Answers are grounded in this paper and cite their sources.
Properly verified and resterilized ICDs appear to be as safe as new devices regarding infection and malfunction rates, offering a safe and cost-effective alternative in resource-limited settings.
AIMS: The safety of pacemaker reuse has been proven by numerous studies in the last two decades. With the exception of one research paper, the safety of reuse of implantable cardioverter-defibrillators has not been properly investigated. Our aim was to establish whether resterilized implantable cardioverter-defibrillators are as safe as new devices in relation to functionality and infection rates. METHODS: All the patients (n = 271) implanted with a new or a donated, used implantable cardioverter-defibrillator (ICD) at the Institute of Cardiovascular Disease Timisoara Romania between January 2001 and December 2012 were included in the study. The patients had class I indication for ICD implantation. One hundred fifty-seven patients received reused ICDs and 114 patients received new ICDs. Complications were defined as infections that required reintervention, device malfunction, and replacements due to untimely or unexpected battery depletion. RESULTS: Complications occurred in 4.38% of the patients in the new ICD group and in 1.91% of the reused ICD group. The difference was not considered statistically significant (odds ratio 0.28, 95% confidence interval 0.04-1.82, P = .18). CONCLUSION: According to our data, properly verified and resterilized ICDs are as safe as new devices, when risk of infection or malfunction rates are assessed. Due to the high costs of new ICDs, their safe reuse has profound humanitarian and financial implications.
No takes yet. Share an insight, caveat, or question.
Enache et al. (2019) studied this question.