Consumption of multicomponent dietary weight loss supplements can trigger atrial tachycardia leading to inappropriate ICD shocks.
Answers are grounded in this paper and cite their sources.
May precipitate inappropriate ICD shocks via supplement-triggered atrial tachycardia; leaves open broader arrhythmogenic risks in device patients.
This report describes the clinical course of a 40-year-old female who experienced repetitive ICD firing after consuming Metabolife 356, a multicomponent dietary weight loss supplement. Following the initiation of Metabolife 356, the patient experienced four shocks over a 3 day period with two 30 J shocks being delivered sequentially. Interrogation of the device revealed atrial tachycardia with 1:1 AV conduction at a rate of 240 beats/min. Metabolife 356 was discontinued and the dosage of sotalol was increased to 120 mg twice daily without recurrence of ICD discharge.
No takes yet. Share an insight, caveat, or question.
McBride et al. (2004) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: