A technique for low-level alpha-particle autoradiography of bronchial epithelium utilising the plastic track detector, CR-39, is described. This plastic is new to the field of nuclear track detection and is very sensitive to alpha-particles. The recording properties of CR-39 for alpha-particles are described in detail and the technique for autoradiography discussed. This technique includes two methods of background reduction enabling activities as low as 2 alpha-particles cm-2 to be detected. The location of the point of emission of an alpha-particle from the tissue surface can be determined to an accuracy of a few micrometers. Determination of the lower limit to the energy of individual alpha-particles is possible from measurements of their range in the plastic. Examples are given of the determination of the microdistribution of alpha-particle active nuclei in bronchial tissue including the observation of two 'hot-spots' in the epithelium sample which are attributed to the presence of small particles of uranium with its daughter products.
No takes yet. Share an insight, caveat, or question.
Henshaw et al. (1979) studied this question.
Synapse has enriched one closely related paper. Consider it for comparative context: