Answers are grounded in this paper and cite their sources.
A method for the absolute calibration of a flat optical surface is described. An algorithm which utilizes the orthogonal properties of the Zernike polynomials was used on interferometric data to fabricate and certify three 10-inch Zerodur reference flats to X/100 peak-to-valley quality.
No takes yet. Share an insight, caveat, or question.
Bernard S. Fritz (1984) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: