Key result
The OFDI catheter-based prediction method yielded a higher median %Correct area compared to the wire-based method when the catheter and wire were in contact.
Why the study?
The study was conducted to determine the predictive accuracy of OFDI on the debulking effects of rotational atherectomy and compare OFDI catheter-based with Rota wire-based prediction methods.
Can optical frequency domain imaging (OFDI) accurately predict the debulking effect of rotational atherectomy in patients with severe calcified coronary lesions?
Observational (n=55)
No
Can optical frequency domain imaging (OFDI) accurately predict the debulking effect of rotational atherectomy in patients with severe calcified coronary lesions?
OFDI-based simulation of rotational atherectomy effects is feasible and its accuracy depends on the relative position of the OFDI catheter and guidewire, potentially aiding in reducing peri-procedural complications.
Answers are grounded in this paper and cite their sources.
OFDI catheter-based prediction may refine RA planning accuracy; leaves open whether it improves outcomes or warrants routine adoption.
This study determined the predictive accuracy of optical frequency domain imaging (OFDI) on debulking effects of rotational atherectomy (RA) and compared the predictive accuracy of OFDI catheter-based with Rota wire-based prediction methods. This prospective, single-center, observational study included 55 consecutive patients who underwent OFDI-guided RA. On pre-RA OFDI images, a circle, identical to the Rota burr was drawn at the center of the OFDI catheter (OFDI catheter-based prediction method) or wire (wire-based prediction method). The area overlapping the vessel wall was defined as the predicted ablation area (P-area). The actual ablated area (A-area) was measured by superimposing the OFDI images before and after RA. The overlapping P-area and A-area were defined as overlapped ablation area (O-area), and the predictive accuracy was evaluated by %Correct area (O-area/P-area) and %Error area (A-area - O-area/A-area). The median %Correct and %Error areas were 47.8% and 41.6%, respectively. Irrelevant ablation (low %Correct-/high % Error areas) and over ablation (high %Correct-/high % Error areas) were related to deep vessel injury and intimal flap outside the P-area. The predictive accuracy was better in the OFDI catheter-based prediction method than the wire-based prediction method in the cross sections where the OFDI catheter and wire came in contact. However, it was better in the latter than the former where the OFDI catheter and wire were not in contact. OFDI-based simulation of the RA effect is feasible though accuracy may be affected by the OFDI catheter and wire position. OFDI-based simulation of RA effect might reduce peri-procedural complications during RA.
No takes yet. Share an insight, caveat, or question.
Hamana et al. (2023) conducted an observational in Severely calcified coronary lesions (n=55). OFDI catheter-based prediction method vs. Wire-based prediction method was evaluated on Median %Correct area in cross sections where OFDI catheter and wire were in contact (p=<0.001). The OFDI catheter-based prediction method yielded a higher median %Correct area compared to the wire-based method when the catheter and wire were in contact.
Synapse has enriched 10 related papers. Consider them for comparative context: