Why the study?
The study was designed to assess the safety and effectiveness of the 17-mm Avalus bioprosthesis in patients undergoing surgical aortic valve replacement.
Is the 17-mm Avalus bioprosthesis safe and effective in patients undergoing surgical aortic valve replacement?
Is the 17-mm Avalus bioprosthesis safe and effective in patients undergoing surgical aortic valve replacement?
The 17-mm Avalus aortic bioprosthesis demonstrates acceptable safety and echocardiographic performance at 2 years in a small Japanese cohort.
Answers are grounded in this paper and cite their sources.
Supports 17-mm Avalus use in small annuli; extends limited RCT data but leaves durability open.
BACKGROUND: The PERIcardial SurGical AOrtic Valve ReplacemeNt (PERIGON) Japan Trial was developed to assess the safety and effectiveness of the 17-mm Avalus bioprosthesis (Medtronic, Minneapolis, MN, USA) in patients undergoing surgical aortic valve replacement. METHODS AND As such, 9 of 11 patients (82%) successfully met the primary endpoint. One death occurred between the 1- and 2-year follow-up visits, unrelated to the valve. There were no valve reinterventions, explants, or device deficiencies through 2 years. CONCLUSIONS: The PERIGON Japan Trial met its primary endpoint. Surgical implantation of the 17-mm Avalus aortic bioprosthesis can be performed with an acceptable incidence of device-related adverse events, and the valve performs effectively based on echocardiographic findings.
No takes yet. Share an insight, caveat, or question.
Okita et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: