Key result
The Avalus stented aortic valve bioprosthesis demonstrated comparable haemodynamic performance to the Magna valve, with a lower left ventricular mass index at 1 year postoperatively.
Why the study?
The Avalus stented bovine pericardial valve is safe and effective in Western populations, but its use in Japanese patients is poorly understood.
Does the Avalus stented aortic valve bioprosthesis provide comparable safety and haemodynamic performance to the Magna valve in Japanese patients undergoing surgical aortic valve replacement?
Cohort (n=468)
No
Does the Avalus stented aortic valve bioprosthesis provide comparable safety and haemodynamic performance to the Magna valve in Japanese patients undergoing surgical aortic valve replacement?
Answers are grounded in this paper and cite their sources.
The Avalus stented aortic valve bioprosthesis demonstrates comparable safety and haemodynamic performance to the established Magna valve in Japanese patients, with potentially greater left ventricular mass regression at 1 year.
OBJECTIVES: A new stented bovine pericardial valve (Avalus™) has been proven safe and effective with good hemodynamic performance in Western populations. However, its use in Japanese patients is poorly understood. We retrospectively compared the feasibility, safety, and valve haemodynamics between the Avalus™ and Magna™ valves in patients who underwent surgical aortic valve replacement (SAVR). METHODS: This study included 87 patients receiving an Avalus™ valve and 387 receiving a Magna™ valve. We evaluated adverse events, outcomes, and valve haemodynamics within 1 year postoperatively. There were no significant differences in any surgical risk scores. RESULTS: No in-hospital mortality occurred in the Avalus™ group, but two mortality events occurred in the Magna™ group. No pacemaker implantation for complete atrioventricular block was required in the Avalus™ group. There was no significant difference in in-hospital or clinical outcomes between the two groups until 1 year postoperatively. Left ventricular mass index reduction appeared to predominate in the Avalus™ over Magna™ group. There was no significant difference in the mean pressure gradient or effective orifice area of each valve size at 1 week or 1 year between the two groups, apart from the mean pressure gradient of the 23-mm valve at 1 week. Three patients (3.4%) in the Avalus™ group and 39 (10.8%) in the Magna™ group (p = 0.12) had severe patient-prosthesis mismatch at 1 week postoperatively. CONCLUSIONS: The new Avalus™ stented aortic valve bioprosthesis was associated with good in-hospital outcomes and good valve functionality post-SAVR in Japanese patients.
No takes yet. Share an insight, caveat, or question.
Tadokoro et al. (2021) conducted a cohort in Severe aortic valve stenosis and insufficiency (n=468). Avalus stented aortic valve bioprosthesis vs. Magna valve was evaluated on Left ventricular mass index (LVMI) at 1 year postoperatively (adjusted difference 11.52, 95% CI 2.79-20.25, p=0.010). The Avalus stented aortic valve bioprosthesis demonstrated comparable haemodynamic performance to the Magna valve, with a lower left ventricular mass index at 1 year postoperatively.
Synapse has enriched 10 related papers. Consider them for comparative context: