Key result
Avalus aortic valve meets standard safety criteria overall but exceeds bleeding thresholds at ~3.7%/valve-year.
Why the study?
Does the Avalus bovine stented aortic valve bioprosthesis achieve acceptable late valve-related adverse event rates (<2x objective performance criteria) and satisfactory haemodynamic performance in patients undergoing surgical aortic valve replacement?
Cohort (n=686)
non-randomized
Yes
Does the Avalus bovine stented aortic valve bioprosthesis achieve acceptable late valve-related adverse event rates (<2x objective performance criteria) and satisfactory haemodynamic performance in patients undergoing surgical aortic valve replacement?
Answers are grounded in this paper and cite their sources.
First multicentre pivotal analysis of the Avalus bovine pericardial surgical aortic bioprosthesis shows favourable 1-year survival (96.4%), haemodynamics, and low valve-related complication rates meeting ISO/FDA objective performance criteria, except for late bleeding rates primarily driven by anticoagulation for non-valvular comorbidities.
OBJECTIVES: We assessed the safety, effectiveness and haemodynamic performance of a new bovine stented aortic valve bioprosthesis (Avalus™). METHODS: The PERIGON Pivotal Trial is a prospective, non-randomized, multicentre study. Subjects had symptomatic moderate or severe aortic stenosis or chronic, severe aortic regurgitation. Death, valve-related adverse events (AEs), functional recovery and haemodynamic performance were assessed at discharge, 3-6 months and 1 year. The primary analysis compared 'late' (>30 days post-implant) linearized rates of valve-related thromboembolism, thrombosis, all and major haemorrhage, all and major paravalvular leak (PVL) and endocarditis after implantation with objective performance criteria (OPC) for AEs, in accordance with EN ISO 5840:2009. We hypothesized that the upper 95% confidence bounds of the true linearized AE rates would be ≥ 2 × OPC; rejection of the null hypothesis would demonstrate that these rates were below acceptable rates. The analysis was required to include at least 150 patients followed to 1 year and 400 valve-years. Kaplan-Meier survival analysis was also performed. RESULTS: Total number of valve-years was 459.5 (n = 686). Linearized rates were <2 × OPC for death and valve-related thromboembolism, valve thrombosis, all and major PVL, and endocarditis, but ≥2 × OPC for all and major haemorrhage. Survival at 1 year (n = 270) was 96.4%. Patients showed good functional recovery, and haemodynamic performance was within expected range. CONCLUSIONS: This analysis demonstrated a good safety profile and clinical effectiveness of the Avalus valve except for bleeding rates. The linearized rates of all and major haemorrhage may be related to long-term anticoagulation for non-valvular indications and the length of follow-up of this cohort. TRIAL REGISTRATION: NCT02088554 (www.clinicaltrials.gov).
No takes yet. Share an insight, caveat, or question.
Klautz et al. (2017) conducted a cohort in Symptomatic moderate or severe aortic stenosis or chronic severe aortic regurgitation (n=686). Avalus bovine stented pericardial aortic valve bioprosthesis vs. Objective performance criteria (OPC) defined by ISO 5840:2009 and FDA guidance was evaluated on Late (>30 days post-implant) linearized rates of valve-related adverse events (thromboembolism, thrombosis, all and major haemorrhage, all and major paravalvular leak, and endocarditis) compared against 2 x objective performance criteria (OPC). The Avalus aortic bioprosthesis achieved 96.4% 1-year survival and late adverse event rates <2x OPC for thromboembolism, PVL, and endocarditis, but exceeded 2x OPC for all and major haemorrhage.