Key result
Transcranial Doppler (TCD) demonstrated higher sensitivity compared to transesophageal echocardiography for detecting right to left shunts.
Why the study?
Does transcranial Doppler (TCD) improve detection of right to left shunts compared to transesophageal echocardiography (TEE) in patients with cryptogenic stroke and transient ischemic attacks?
Observational (n=118)
Does transcranial Doppler (TCD) improve detection of right to left shunts compared to transesophageal echocardiography (TEE) in patients with cryptogenic stroke and transient ischemic attacks?
TCD demonstrated higher sensitivity than TEE for detecting right to left shunts, highlighting the need for standardized performance protocols for TEE to reduce false negatives and interoperator variability.
Answers are grounded in this paper and cite their sources.
TCD may detect more right-to-left shunts than TEE in cryptogenic stroke; leaves open need for TEE standardization and outcome trials before practice change.
BACKGROUND AND PURPOSE: Although transesophageal echocardiography (TEE) is the gold standard for right to left shunt detection, we observed that transcranial Doppler (TCD) was more sensitive and sought an explanation. METHODS: We retrospectively evaluated results of TCD and TEE in 118 patients with cryptogenic stroke and transient ischemic attacks. TCDs were done as per modification of a published performance protocol and interpreted by a neurologist and radiologist. TEEs were performed and interpreted by five cardiologists without standardized protocol. Statistical methodology included χ(2) tests, Fisher exact tests, and ANOVA. RESULTS: Overall agreement between TCD and TEE was found for 76 of 118 patients. Sensitivities of TCD and TEE were 93.8% and 53.1%. Sensitivities for TCD interpreters were 61.1% and 64.1%. Sensitivities for TEE operators varied from 46.7% to 75.7%. We were unable to find a performance protocol for TEE microbubble studies in either published literature or certification organizations. CONCLUSIONS: TEEs and TCDs are operator dependent and thus subject to false negatives. The lower yield and interoperator variability in TEE results appear to reflect the lack of performance protocols and engender concern about false negatives in community use. Consensus performance protocols and certification criteria for both modalities should have an impact on accuracy of shunt detection.
No takes yet. Share an insight, caveat, or question.
Chang et al. (2012) conducted an observational in Cryptogenic stroke and transient ischemic attacks (n=118). Transcranial Doppler (TCD) vs. Transesophageal echocardiography (TEE) was evaluated on Sensitivity for right to left shunt detection. Transcranial Doppler (TCD) demonstrated higher sensitivity compared to transesophageal echocardiography for detecting right to left shunts.
Synapse has enriched 10 related papers. Consider them for comparative context: