Key result
Multiple relative AV delay optimization using invasive LV dp/dt max had better reproducibility (SDD 23ms, R2=0.76) than absolute methods (p<0.01).
Why the study?
Does a multiple relative measurement protocol improve the reproducibility of AV delay optimization in patients with CRT compared to absolute measurements?
Does a multiple relative measurement protocol improve the reproducibility of AV delay optimization in patients with CRT compared to absolute measurements?
p-value: p=<0.01
AV delay optimization using multiple relative measurements improves reproducibility compared to absolute measurements and doubles the hemodynamic increment of CRT compared to nominal AV delay.
May support relative AV optimization in CRT; hypothesis-generating without outcome data.
Background Reproducibility and hemodynamic efficacy of optimization of AV delay (AVD) of cardiac resynchronization therapy (CRT) using invasive LV dp/dt max are unknown. Method and results 25 patients underwent AV delay (AVD) optimisation twice, using continuous left ventricular (LV) dp/dt max , systolic blood pressure (SBP) and pulse pressure (PP). We compared 4 protocols for comparing dp/dt max between AV delays: Immediate absolute: mean of 10s recording of dp/dt max acquired immediately after programming the tested AVD, Delayed absolute: mean of 10s recording acquired 30s after programming AVD, Single relative: relative difference between reference AVD and the tested AVD, Multiple relative: averaged difference, from multiple alternations between reference and tested AVD. We assessed for dp/dt max , LVSBP and LVPP, test–retest reproducibility of the optimum. Optimization using immediate absolute dp/dt max had poor reproducibility (SDD of replicate optima=41ms; R 2 =0.45) as did delayed absolute (SDD 39ms; R 2 =0.50). Multiple relative had better reproducibility: SDD 23ms, R 2 =0.76, and (p<0.01 by F test). Compared with AAI pacing, the hemodynamic increment from CRT, with the nominal AV delay was LVSBP 2% and LVdp/dt max 5%, while CRT with pre-determined optimal AVD gave 6% and 9% respectively. Conclusions Because of inevitable background fluctuations, optimization by absolute dp/dt max has poor same-day reproducibility, unsuitable for clinical or research purposes. Reproducibility is improved by comparing to a reference AVD and making multiple consecutive measurements. More than 6 measurements would be required for even more precise optimization — and might be advisable for future study designs. With optimal AVD, instead of nominal, the hemodynamic increment of CRT is approximately doubled.
No takes yet. Share an insight, caveat, or question.
Whinnett et al. (2013) studied Cardiac resynchronization therapy (n=25). Multiple relative AV delay optimization vs. Absolute AV delay optimization was evaluated on Test-retest reproducibility of the optimum (p=<0.01). Multiple relative AV delay optimization using invasive LV dp/dt max had better reproducibility (SDD 23ms, R2=0.76) than absolute methods (p<0.01).