A rapid HPLC-UV method allows for accurate and simultaneous monitoring of amiodarone and its active metabolite in serum or plasma, potentially improving patient safety.
Answers are grounded in this paper and cite their sources.
May support amiodarone TDM in clinical labs; extends validated HPLC-UV methods for simultaneous DEA quantitation.
This article describes a rapid isocratic high-performance liquid chromatographic (HPLC) method for the simultaneous measurement of the antiarrhythmic drug amiodarone and its potentially active metabolite N-desethylamiodarone (DEA). Following a simple liquid-liquid extraction, amiodarone and its metabolite are quantitated (0.3-6.0 mg/L) by analysis on an HPLC-UV system. The analytical time was reduced by 50%, without compromising the assay performance, when Rocket column technology was employed. The assay's limit of quantitation, linearity, imprecision, and accuracy adequately covered the therapeutic range for appropriate patient monitoring. Amiodarone and DEA can be simultaneously and accurately quantitated in serum or plasma by HPLC-UV detection with imprecision < 6% at therapeutic concentrations and a quantitation range from 0.3 to 6.0 mg/L. Monitoring of this drug can allow for effective use, while minimizing serious side effects.
No takes yet. Share an insight, caveat, or question.
Juenke et al. (2004) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: