Key result
The eSTICO method for noninvasive assessment of cardiac output during exercise showed no significant difference compared to open circuit measures at peak exercise.
Why the study?
Assessing cardiovascular hemodynamics during exercise provides clinical insights but remains challenging to acquire in clinical settings.
Does the eSTICO method accurately measure cardiac output during exercise in healthy adults compared to established methods?
Cross-Sectional (n=14)
Does the eSTICO method accurately measure cardiac output during exercise in healthy adults compared to established methods?
A novel noninvasive method using systolic timing intervals (eSTICO) provides a feasible and relatively accurate assessment of cardiac output during exercise in healthy adults.
Answers are grounded in this paper and cite their sources.
eSTICO may enable simpler exercise hemodynamics assessment; leaves open accuracy and clinical utility pending validation.
Background: Although assessment of cardiovascular hemodynamics during exercise can provide clinical insights, it is challenging to acquire it in clinical settings. Objectives: Accordingly, this preliminary study was to determine whether a novel elaboration on systolic time interval measures (eSTICO) method of quantifying cardiac output and stroke volume was comparable to those obtained using a validated soluble gas (open circuit CO measure [OpCircCO]) method or calculation based on oxygen consumption (oxygen consumption-based CO [VO₂CO]) during exercise. Methods: For the present study, 14 healthy subjects (male: n = 12, female: n = 2) performed incremental exercise on a recumbent cycle ergometer. At rest and during exercise, cardiac output (CO) was obtained via the eSTICO method, while the OpenCircCO and VO₂CO measures were obtained at the last minute of each workload. Results: At peak, there was no difference between eSTICO and OpCircCO (12.39 ± 3.06 vs. 13.96 ± 2.47 L/min, p > 0.05), while there was a slight difference between eSTICO and VO₂CO (12.39 ± 3.06 vs. 14.28 ± 2.55 L/min, p < 0.05). When we performed correlation analysis with all subjects and all measures of CO at all WL, between eSTICO and OpenCircCO, there was a good relationship (r = 0.707, p < 0.001) with a Bland and Altman agreement analysis demonstrating a −1.6 difference (95% LoA: −6.3–3.5). Between eSTICO and VO₂CO, we observed an r = 0.865 (p < 0.001) and a Bland and Altman agreement analysis with a −1.2 difference (95% LoA: −4.8–2.4). Conclusion: A novel exploitation of cardiac hemodynamics using systolic timing intervals may allow a relatively good assessment of CO during exercise in healthy adults.
No takes yet. Share an insight, caveat, or question.
Kim et al. (2023) conducted a cross-sectional in Healthy adults (n=14). eSTICO method vs. Open circuit CO measure (OpCircCO) and oxygen consumption-based CO (VO2CO) was evaluated on Cardiac output at peak exercise (MD -1.6, 95% CI -6.3-3.5, p=> 0.05). The eSTICO method for noninvasive assessment of cardiac output during exercise showed no significant difference compared to open circuit measures at peak exercise.