Key result
Jacket-based ECG collection (JET) detected significant QTc prolongation of 27, 54, and 76 msec after 8, 16, and 32 mg/kg sotalol, correlating well with implanted telemetry.
Why the study?
Does a non-invasive jacket-based external telemetry system accurately detect sotalol-induced QTc prolongation compared to implanted telemetry in conscious non-human primates?
Does a non-invasive jacket-based external telemetry system accurately detect sotalol-induced QTc prolongation compared to implanted telemetry in conscious non-human primates?
Jacket-based external telemetry is a reliable, non-invasive alternative to implanted telemetry for detecting drug-induced QTc prolongation in conscious non-human primates.
Supports JET for non-invasive QTc monitoring in NHPs; extends telemetry validation but leaves human translation open.
BACKGROUND AND PURPOSE: During repeat-dose toxicity studies, ECGs are collected from chemically or physically-restrained animals over a short timeframe. This is problematic due to cardiovascular changes caused by manual restraint stress and anesthesia, and limited ECG sampling. These factors confound data interpretation, but may be overcome by using a non-invasive jacket-based ECG collection (JET). The current study investigated whether a jacketed external telemetry system could detect changes in cardiac intervals and heart rate in non-human primates (NHPs), previously implanted with a PCT transmitter. EXPERIMENTAL APPROACH: Twelve male cynomolgus monkeys were treated weekly with vehicle or sotalol (8, 16, 32 mg kg⁻¹) p.o. ECGs were collected continuously for 24 hours, following treatment, over 4 weeks. A satellite group of six NHPs was used for sotalol toxicokinetics. KEY RESULTS: Sotalol attained Cmax values 1-3 hours after dosing, and exhibited dose-proportional exposure. In jacketed NHPs, sotalol dose-dependently increased QT/QTc intervals, prolonged PR interval, and reduced heart rate. Significant QTc prolongation of 27, 54 and 76 msec was detected by JET after 8, 16, and 32 mg kg⁻¹ sotalol, respectively, compared with time-matched vehicle-treated animals. Overall, JET-derived PR, QT, QTc intervals, QRS duration, and heart rate correlated well with those derived from PCT. CONCLUSIONS AND IMPLICATIONS: The current findings clearly support the use of JET to quantify cardiac interval and rhythm changes, capable of detecting QTc prolongation caused by sotalol. JET may be a preferred method compared to restraint-based ECG because high-density ECG sampling can be collected in unstressed conscious monkeys, over several weeks.
No takes yet. Share an insight, caveat, or question.
Derakhchan et al. (2013) studied QTc interval prolongation (n=18). Jacket-based ECG collection (JET) vs. Implanted telemetry (PCT) and vehicle was evaluated on QTc prolongation. Jacket-based ECG collection (JET) detected significant QTc prolongation of 27, 54, and 76 msec after 8, 16, and 32 mg/kg sotalol, correlating well with implanted telemetry.