Why the study?
This study aimed to assess the agreement between direct intra-arterial pressure and ClearSight finger cuff arterial pressure measurements in elderly patients undergoing TAVR.
Does ClearSight finger cuff arterial pressure measurement agree with direct intra-arterial pressure in elderly patients undergoing TAVR?
Population
30 patients aged 65 years and older undergoing TAVR under general anesthesia in Japan
Comparison
Direct intra-arterial pressure vs ClearSight finger cuff arterial pressure measurements
Design
Prospective observational study
Follow-up
30 min after valve deployment
Key result
ClearSight finger cuff arterial pressure measurement showed acceptable agreement with direct intra-arterial pressure in elderly TAVR patients, with a systolic blood pressure bias of -4.88 mmHg.
Authors
Loading...
ClearSight shows good agreement with intra-arterial BP in TAVR; leaves open non-invasive monitoring as alternative pending larger validation.
Observational (n=30)
No
Does ClearSight finger cuff arterial pressure measurement agree with direct intra-arterial pressure in elderly patients undergoing TAVR?
Effect estimate: Bias -4.88 ± 15.46 mmHg for SBP, 4.73 ± 8.95 mmHg for mean, and 9.53 ± 9.01 mmHg for DBP
Non-invasive ClearSight finger cuff arterial pressure monitoring provides acceptable agreement with direct intra-arterial pressure in elderly patients undergoing TAVR, offering a potential alternative for hemodynamic monitoring.
Yahagi et al. (2023) conducted an observational in Elderly patients undergoing transcatheter aortic valve replacement (TAVR) (n=30). ClearSight finger cuff arterial pressure measurement vs. Direct intra-arterial pressure measurement was evaluated on Agreement between direct intra-arterial pressure and ClearSight finger cuff arterial pressure measurements (Bias -4.88 ± 15.46 mmHg for SBP, 4.73 ± 8.95 mmHg for mean, and 9.53 ± 9.01 mmHg for DBP). ClearSight finger cuff arterial pressure measurement showed acceptable agreement with direct intra-arterial pressure in elderly TAVR patients, with a systolic blood pressure bias of -4.88 mmHg.