Key result
EndoPAT measurement did not detect decreased endothelial function in renally impaired or diabetic patients compared to healthy volunteers.
Why the study?
Is the EndoPAT a feasible and discriminating tool to assess endothelial function and its acute changes in different populations?
Observational
Is the EndoPAT a feasible and discriminating tool to assess endothelial function and its acute changes in different populations?
The EndoPAT device may not be suitable for assessing acute changes in endothelial function in early-phase clinical pharmacology studies, as it failed to discriminate between healthy and diseased populations or detect effects of robust interventions.
Answers are grounded in this paper and cite their sources.
EndoPAT may lack sensitivity for endothelial dysfunction in renally impaired or diabetic patients; leaves open its utility and requires validation before adoption.
Endothelial dysfunction is a potential target for (pharmaceutical) intervention of several systemic pathological conditions. We investigated the feasibility of the EndoPAT to evaluate acute changes in endothelial function with repeated noninvasive measurements and assessed its discriminating power in different populations. Endothelial function was stable over a longer period of time in renally impaired patients (coefficient of variation 13%). Endothelial function in renally impaired and type 2 diabetic patients was not decreased compared to healthy volunteers (2.9 ± 1.4 and 1.8 ± 0.3, resp., versus 1.8 ± 0.5, P > 0.05). The EndoPAT did not detect an effect of robust interventions on endothelial function in healthy volunteers (glucose load: change from baseline 0.08 ± 0.50, 95% confidence interval -0.44 to 0.60; smoking: change from baseline 0.49 ± 0.92, 95% confidence interval -0.47 to 1.46). This suggests that at present the EndoPAT might not be suitable to assess (changes in) endothelial function in early-phase clinical pharmacology studies. Endothelial function as measured by the EndoPAT could be physiologically different from endothelial function as measured by conventional techniques. This should be investigated carefully before the EndoPAT can be considered a useful tool in drug development or clinical practice.
No takes yet. Share an insight, caveat, or question.
Moerland et al. (2012) conducted an observational in Endothelial dysfunction. EndoPAT measurement vs. Healthy volunteers was evaluated on Endothelial function (p=>0.05). EndoPAT measurement did not detect decreased endothelial function in renally impaired or diabetic patients compared to healthy volunteers.
Synapse has enriched 10 related papers. Consider them for comparative context: