Key result
Real time simultaneous triplane echocardiography (RT3P) demonstrated a significantly better inverse correlation between functional area and mean gradient than the Doppler continuity equation (P=0.0359).
Why the study?
Does real time triplane echocardiography improve the assessment of the functional area of prosthetic aortic valves compared to the Doppler continuity equation in patients requiring follow up?
Observational (n=23)
Does real time triplane echocardiography improve the assessment of the functional area of prosthetic aortic valves compared to the Doppler continuity equation in patients requiring follow up?
p-value: p=0.0359
Real time triplane echocardiography is a feasible, reliable, and faster alternative to the standard Doppler continuity equation for assessing prosthetic aortic valve area.
Should not yet change prosthetic aortic valve assessment practices; leaves open whether RT3P improves outcomes over Doppler continuity equation.
PURPOSE: Our study is aimed at evaluating the feasibility and reliability of a simple method for the measurement of the functional area of prosthetic aortic valves (EOA). Three-dimensional echocardiography has proven accurate for left ventricular volume, stroke volume, and aortic valve area measurement. We studied the feasibility and reliability of real time simultaneous triplane echocardiography (RT3P) for assessing the EOA with a fast formula based on the principle of continuity equation, in which we replaced Doppler-derived stroke volume (SV) with SV directly measured with RT3P. METHODS AND RESULTS: EOA of prosthetic aortic valves were measured in 23 consecutive patients requiring periodical follow up. EOA was calculated using Doppler continuity equation (DCE) and the RT3P method by replacing Doppler-derived SV with SV measured with real time triplane echocardiography. We compared functional areas obtained with the two methods with the prosthetic area indicated in the manufacturer's specifications and with the mean transprosthetic gradient. Both methods had a good correlation with the area indicated by the manufacturer. RT3P revealed an inverse correlation between functional area and mean gradient that was better than DCE (P = 0.0359). Inter- and intraobserver variability was not different between the two methods. Execution time was significantly shorter for RT3P. CONCLUSIONS: RT3P is a simple method that can be performed quite rapidly, and can complement the overall assessment of prosthetic valve function. Further studies can confirm our technique.
No takes yet. Share an insight, caveat, or question.
Alunni et al. (2011) conducted an observational in Prosthetic aortic valves (n=23). Real time simultaneous triplane echocardiography (RT3P) vs. Doppler continuity equation (DCE) was evaluated on Correlation between functional area and mean transprosthetic gradient (p=0.0359). Real time simultaneous triplane echocardiography (RT3P) demonstrated a significantly better inverse correlation between functional area and mean gradient than the Doppler continuity equation (P=0.0359).