Outcome and safety of intraoperative defibrillation testing during device replacement: the Simpler trial

View Full Paper
Ask AI
Bookmark
Share

Why the study?

Older registries showed an increased incidence of significant findings during defibrillation testing at replacement, leaving its frequency of significant findings and safety uncertain.

Does intraoperative defibrillation testing provide clinically important information or cause complications in patients undergoing ICD generator replacement?

Population

92 patients undergoing ICD generator replacement

Comparison

Defibrillation testing during ICD generator replacement

Design

Prospective observational multi-centre study

Follow-up

6-months post-procedure

Authors

AMAnat MilmanNational Archives and Records AdministrationENEyal NofElectrophysiologyMRMoshe RavElectrophysiology

Discussion

Loading...

Member takes

Implication

DT during ICD replacement appears safe with low yield; leaves open whether routine testing should be abandoned pending larger studies.

Structured PICO

Does intraoperative defibrillation testing provide clinically important information or cause complications in patients undergoing ICD generator replacement?

P
Population
92 patients undergoing ICD generator replacement, median age 68 years, 79.8% male.
I
Intervention
Intraoperative defibrillation testing (DT) during ICD generator replacement
O
Outcome
Failure to terminate induced ventricular fibrillation (VF) with a single shock 10 J below the maximal capacity of the devicesurrogate

Intraoperative defibrillation testing during ICD generator replacement is safe but does not yield clinically important information, supporting the trend to abandon this practice.

Cite This Study

Milman et al. (2023) studied this question.

synapsesocial.com/papers/6a7cb41498c0e0ecb2770e01https://doi.org/10.1093/europace/euac282

Topics

Cardiac device therapy
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Necessity for Surgical Revision of Defibrillator Leads Implanted Long-Term2008 · 150 citations
  2. 2Incidence and Costs Related to Lead Damage Occurring Within the First Year After a Cardiac Implantable Electronic Device Replacement Procedure2016 · 34 citations
  3. 3Amiodarone or an Implantable Cardioverter–Defibrillator for Congestive Heart Failure2005 · 6,587 citations
  4. 4Failure to Deliver Therapy by a Riata Lead with Internal Wire Externalization and Normal Electrical Parameters During Routine Interrogation2012 · 25 citations
  5. 5Intra-operative defibrillation testing and clinical shock efficacy in patients with implantable cardioverter-defibrillators: the NORDIC ICD randomized clinical trial2015 · 118 citations