Highlights coronary sinus perforation as a rare but life-threatening complication of left ventricular pacing via the coronary sinus, which is sometimes utilized after tricuspid valve replacement.
Answers are grounded in this paper and cite their sources.
Warrants long-term vigilance for delayed, life-threatening complications of left ventricular pacing;.
In the presence of tricuspid valve intervention, implantation of the right ventricle lead can be hazardous for tricuspid valve malfunction and can lead to regurgitation. Few cases have been reported with successful left ventricular pacing via the coronary sinus (CS) after tricuspid valve replacement. We represent a 43 year old female patient who had implanted a DDD-R permanent pacemaker via the CS due to symptomatic bradyarrhythmia soon after triple (aortic, mitral and tricuspid) valve stenosis 6 years ago (Fig. 1). We aimed to highlight an uncommon and life threating complication lately diagnosed as CS perforation caused by the CS pacemaker lead (Fig. 2). Figure options Download full-size image Download high-quality image (132 K) Download as PowerPoint slide
No takes yet. Share an insight, caveat, or question.
Soydan et al. (2017) studied this question.