In patients with a prosthetic tricuspid valve, pacing in the middle cardiac vein is a viable alternative to open-chest epimyocardial lead placement.
Answers are grounded in this paper and cite their sources.
May support middle cardiac vein pacing in prosthetic tricuspid valve patients; hypothesis-generating and requires prospective confirmation.
A man with acute endocarditis developed complete heart block several days after the tricuspid and aortic valve replacement. Several weeks after implantation, his epimyocardial pacing leads developed a high threshold and failed to capture the ventricle at the maximal pulse width and output of the pacemaker. An angled-tip lead was placed in the middle cardiac vein for ventricular pacing. The pacing and sensing thresholds of this lead were within the expected range during follow-up. Therefore, in patients with prosthetic tricuspid valve, pacing in the middle cardiac vein should be considered before open-chest placement of the epimyocardial lead.