Why the study?
Does a local verification process accurately assign ISI and MNPT for unsupported reagent/instrument combinations compared to a reference method?
Population
25 random normal donor samples and 13 warfarin pools created from stable warfarin patient plasma
Comparison
RecombiPlasTin 2G evaluated on unsupported… vs ACL TOP 700 (reference method)
Design
Other
Authors
Loading...
Local verification of ISI/MNPT may ensure accurate INR on unsupported pairs; leaves open standardized calibration protocols.
Does a local verification process accurately assign ISI and MNPT for unsupported reagent/instrument combinations compared to a reference method?
A local verification process using normal donor samples and warfarin pools can accurately assign ISI and MNPT values for unsupported reagent/instrument combinations, ensuring accurate INR calculation.
Tange et al. (2013) studied this question.