Answers are grounded in this paper and cite their sources.
This letter introduces a method for power density (PD) measurement of low-power millimeter-wave devices, accounting for antenna/body coupling. This technique employs a thin solid absorbing structure with equivalent scattering properties used to convert the absorbed power into a heat pattern measured by an infrared (IR) camera. The measured IR pattern is then used to reconstruct the PD distribution. The lock-in technique is used to enhance the signal-to-noise ratio. It consists in filtering the ambient IR noise as well as removing the parasitic heat conduction effect. The reduction of noise enables achieving a measurement sensitivity of the order of 1 mW/cm², substantially overcoming the sensitivity without lock-in ($> $10 dB). The proposed approach is experimentally validated for a conical horn antenna at 60 GHz. For the first time, this study demonstrates the sensitivity of IR-based measurements sufficient enough to assess the compliance of medium- and low-power wireless devices above 6 GHz.
No takes yet. Share an insight, caveat, or question.
Ziane et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: