Why the study?
Does a portable pupillometer using Polaroid photography accurately measure pupillary abnormality in diabetic autonomic neuropathy?
Does a portable pupillometer using Polaroid photography accurately measure pupillary abnormality in diabetic autonomic neuropathy?
A simple portable pupillometer using Polaroid photography provides a reliable and repeatable method for diagnosing pupillary abnormalities in diabetic autonomic neuropathy.
Answers are grounded in this paper and cite their sources.
Portable pupillometry may aid bedside detection of diabetic autonomic neuropathy; hypothesis-generating pending prospective validation.
A simple portable pupillometer using Polaroid photography has been developed for use in the diagnosis of diabetic autonomic neuropathy. The pupil diameter per cent (PD%), defined as the percentage of iris diameter to which pupil diameter had dilated in darkness, was measured from the instant colour prints. An age-related normal range was constructed from measurements in 163 healthy subjects in whom PD% declined linearly by 3% per decade. The PD% was measured in 15 patients with automatic neuropathy and 12 were identified with abnormally low values indicating a failure of sympathetically mediated dilatation. Results were consistent with the standard television pupillometric technique and had the same high repeatability with a coefficient of variation of 3.2%.
No takes yet. Share an insight, caveat, or question.
Smith et al. (1986) studied this question.