Why the study?
Does continuous acquisition MRI accurately measure time-averaged flow in the carotid arteries compared to Doppler ultrasound in normal volunteers?
Does continuous acquisition MRI accurately measure time-averaged flow in the carotid arteries compared to Doppler ultrasound in normal volunteers?
Answers are grounded in this paper and cite their sources.
Continuous acquisition MRI provides a valid and reproducible non-invasive method for measuring time-averaged flow, comparable to Doppler ultrasound, particularly useful for vessels inaccessible to ultrasound.
A continuous acquisition method was used to measure the time-averaged flow in the carotid arteries of 10 normal volunteers, using an interleaved flow-sensitive and flow-compensated field echo sequence on a 1.5 T magnetic resonance (MR) system. Validation of the sequence and technique was performed using a pulsatile rotating phantom. The measured flow rates for the common carotid arteries were compared with Doppler ultrasound values obtained immediately after the MR measurement using a Duplex scanner. The correlation (r = 0.52, P less than 0.01) was significant with data spread accounted for by the inherent errors of both techniques. The difference between time-averaged flow measured by MR and Doppler ultrasound was 1.9%. Short-term reproducibility of each technique was assessed by consecutive measurements with values of 6.8% and 6.6% respectively for MR and Doppler; measurements after a 2 - 4 week interval gave a long-term reproducibility of 11.8% and 9.8% respectively. The advantages of continuous acquisition make the method suitable for non-invasive flow measurements, particularly for vessels that are not accessible to ultrasound.
No takes yet. Share an insight, caveat, or question.
Tarnawski et al. (1990) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: