Why the study?
Do different polarity and electrode designs impact acute pacing and sensing characteristics of pacemaker leads in patients undergoing pacemaker implantation?
Do different polarity and electrode designs impact acute pacing and sensing characteristics of pacemaker leads in patients undergoing pacemaker implantation?
Different pacemaker lead designs and polarities do not significantly affect acute pacing and sensing thresholds, though active fixation leads offer placement advantages.
Answers are grounded in this paper and cite their sources.
Supports lead selection flexibility by acute metrics; leaves open long-term outcome comparisons across designs.
In order to analyze the impact of different polarity and electrode designs on the acute pacing and sensing characteristics of pacemaker leads, 80 patients with complete heart block or sinus node dysfunction undergoing pacemaker implantation received eight different leads from five manufacturers. Once the leads were positioned, volt and current thresholds, P wave, peak-to-peak, and slew rate were assessed prospectively. There was no statistically significant difference between acute pacing thresholds, sensing characteristics, or unipolar and bipolar pacing thresholds. Active fixation leads allow atrial mapping and lead placement in areas generally inaccessible to tined tip leads. This is an advantage, especially for patients with a history of open heart surgery.
No takes yet. Share an insight, caveat, or question.
Masterson et al. (1990) studied this question.