Key result
The CineECG-derived ST-T path direction Yh was independently associated with an acute increase in dP/dt max during CRT (regression coefficient 639.4; 95% CI 187.9-1090.9; p=0.006).
Why the study?
A variable proportion of non-responders to CRT warrants new approaches to optimize LV lead position and device programming.
Do CineECG-derived repolarization parameters predict acute hemodynamic response in CRT patients with nonischemic dilated cardiomyopathy and LBBB?
Observational (n=15)
Do CineECG-derived repolarization parameters predict acute hemodynamic response in CRT patients with nonischemic dilated cardiomyopathy and LBBB?
Effect estimate: regression coefficient 639.4 (95% CI 187.9-1090.9)
p-value: p=0.006
CineECG-derived repolarization parameters, specifically the ST-T path direction Yh, can predict acute hemodynamic response to cardiac resynchronization therapy, offering a potential new approach for optimizing LV lead position.
CineECG ST-T direction may predict acute CRT response; hypothesis-generating for repolarization-guided optimization in nonischemic LBBB.
Introduction A variable proportion of non‐responders to cardiac resynchronization therapy (CRT) warrants the search for new approaches to optimize the position of the left ventricular (LV) lead and the CRT device programming. CineECG is a novel ECG modality proposed for the spatial visualization and quantification of myocardial depolarization and repolarization sequences. Objective The present study aimed to evaluate CineECG‐derived parameters in different pacing modes and to test their associations with acute hemodynamic responses in CRT patients. Methods and Results CineECG was used to construct the average electrical path within the cardiac anatomy from the 12‐lead ECG. CineECG and LV dP/dt max were tested in 15 patients with nonischemic dilated cardiomyopathy and left bundle branch block (QRS: 170 ± 17 ms; LVEF: 26 ± 5.5%) under pacing protocols with different LV lead localizations. The CineECG‐derived path directions were computed for the QRS and ST‐T intervals for the anteroposterior (X h ), interventricular (Y h ), and apicobasal (Z h ) axes. In a multivariate linear regression analysis with adjustment for the pacing protocol type, the ST‐T path direction Y h was independently associated with the increase in dP/dt max during CRT, [regression coefficient 639.4 (95% confidence interval: 187.9–1090.9), p = 0.006]. In ROC curve analysis, the ST‐T path direction Y h was associated with the achievement of a 10% increase in dP/dt max (AUC: 0.779, p = 0.002) with the optimal cut‐off > 0.084 (left‐to‐right direction) with sensitivity 0.67 and specificity 0.92. Conclusion The acute hemodynamic response in CRT patients was associated with specific CineECG repolarization sequence parameters, warranting their further testing as potential predictors of clinical outcomes.
No takes yet. Share an insight, caveat, or question.
Sedova et al. (2024) conducted an observational in Nonischemic dilated cardiomyopathy and left bundle branch block (n=15). CineECG-derived ST-T path direction Yh was evaluated on Increase in dP/dt max during CRT (regression coefficient 639.4, 95% CI 187.9-1090.9, p=0.006). The CineECG-derived ST-T path direction Yh was independently associated with an acute increase in dP/dt max during CRT (regression coefficient 639.4; 95% CI 187.9-1090.9; p=0.006).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: