Answers are grounded in this paper and cite their sources.
A procedure has been developed for the analysis of biological materials by inductively coupled plasma mass spectrometry (ICP-MS). Fast, efficient and complete sample digestion is achieved by a combined microwave-nitric acid/open beaker-nitric acid-hydrogen peroxide procedure. The ICP-MS analysis is performed with an on-line five-element internal standard to correct for matrix and instrumental drift effects. Results are presented for 24 elements in three biological reference materials (National Institute of Standards and Technology Standard Reference Materials 5277a Liver and 1566 Oyster and International Atomic Energy Agency Certified Reference Material H4 Animal Muscle). For all elements significantly above the detection limit and reagent blank concentrations, good agreement exists between ICP-MS and certified values.
No takes yet. Share an insight, caveat, or question.
Friel et al. (1990) studied this question.